Zusammenfassung der Ressource
Frage 1
Frage
ABG Analyzers
Directly measures [blank_start]PCO2[blank_end] - Severinghaus, [blank_start]PO2[blank_end] - Clark, pH - [blank_start]Sanz[blank_end]
HCO3, SaO2 and other values are calculated
Formula for read out: (Barometric Pressure - [blank_start]47[blank_end]) x Fractional Concentration of [blank_start]gas[blank_end]
Frage 2
Frage
Severinhaus Electrode
Measures [blank_start]PCO2[blank_end]
[blank_start]Glass[blank_end] electrode in [blank_start]lucite[blank_end] jacket
Includes nylon spacer and [blank_start]teflon[blank_end] membrane
Frage 3
Frage
PCO2 Calibration:
Cylinders containing known concentrations of gas:
Low: [blank_start]5[blank_end]% balance
High: [blank_start]10[blank_end]% slope
Frage 4
Frage
Clark Electrode
Measures [blank_start]PO2[blank_end]
Platinum cathode, silver-silver chloride anode, polypropolene membrane (gas [blank_start]permeable[blank_end])
Using [blank_start]mylar[blank_end] membrane results in insufficient O2 diffusion
Cleaning top with [blank_start]pumice[blank_end] periodically to clean [blank_start]protein[blank_end]
Antworten
-
PO2
-
permeable
-
mylar
-
pumice
-
protein
Frage 5
Frage
PO2 Calibration:
Cylinders containing known concentrations of gas:
Low: [blank_start]0[blank_end]% balance
High: [blank_start]12[blank_end]% or [blank_start]20[blank_end]% slope
Frage 6
Frage
Sanz Electrode
Measures PH
Reference Electrode Contains: [blank_start]KCL[blank_end] solution with known PH, has contsant [blank_start]electrical[blank_end] potential.
Measuring electrode contains: [blank_start]Buffer[blank_end] Solution, silver-silver chloride wire and pH sensitive glass capillary tube
Introduction of sample causes electrical [blank_start]potential[blank_end] to develop. Voltage difference is proportional to PH [blank_start]difference[blank_end]
Antworten
-
KCL
-
electrical
-
Buffer
-
potential
-
difference
Frage 7
Frage
Sanz Electrode Calibration:
Low pH solution: [blank_start]6.84[blank_end] - If displayed value is different than the known sample, adjust with the [blank_start]balance[blank_end] to known value.
High pH solution: [blank_start]7.384[blank_end] - If displayed value is different than the known sample, adjust with the [blank_start]slope[blank_end] control to known value.
Frage 8
Frage
Tonomotry
Used for quality [blank_start]control[blank_end] with ABG analyzers
Allows precise gas mixtures to be [blank_start]equilibrated[blank_end] with buffer solution or whole blood samples
Best method for precise control of [blank_start]Clark[blank_end] electrode
Vert complex - not widely used
Antworten
-
control
-
Clark
-
equilibrated