Chapter 10: Point Dose Calculations

Beschreibung

System of Dosimetric Calculations
Caity Bernier
Karteikarten von Caity Bernier, aktualisiert more than 1 year ago
Caity Bernier
Erstellt von Caity Bernier vor mehr als 7 Jahre
12
0

Zusammenfassung der Ressource

Frage Antworten
Monitor Units (MUs) Machine output; it is calibrated to give 1 cGy/MU at a certain point (ex: dmax along CAX) under certain conditions (ex: 10x10 FS, 100 cm SSD or SAD).
Dose Energy deposited per unit mass. A point quantity
MUs and Dose Different points inside the tissue, along the CAX, receive different amounts of dose with the same Muscle delivered. The dose at a point along the CAX depends on: 1) distance from the source (ISL) 2) depth (attenuation, energy, absorber dependent) 3) scatter (energy and absorber dependent)
MU Calculation (SSD Technique)
MU Calculation (SAD Technique)