Question 1
Question
The x-ray intensity is lower on the anode side of the tube because of the:
Answer
-
Line Focus Principle
-
Heel Effect
-
Focusing Cup
-
Filament Length
Question 2
Question
Which of the following is (are) true regarding the protective housing of the x-ray tube?
Answer
-
Provides mechanical support for the x-ray tube.
-
Protects the tube from damage.
-
Aids in absorbing low energy radiation created within the x-ray tube.
-
All of the above
Question 3
Question
The negative side of the x-ray tube holds the:
Answer
-
Filament
-
Anode
-
Target
-
Rotor
Question 4
Question
Tube arcing is caused by:
Question 5
Question
A dual focus tube has two:
Answer
-
Anodes
-
Filaments
-
Rotors
-
Stators
Question 6
Question
An x-ray tube has a tube rating chart to show:
Answer
-
Housing cooling times
-
Anode cooling times
-
Maximum exposure times
-
Maximum patient doses
Question 7
Question
This device is located outside the glass envelope:
Answer
-
Stator
-
Rotor
-
Focal Spot
-
Filament
Question 8
Question
Which of the following is needed to produce an x-ray?
Question 9
Question
The tube current is changed by changing the:
Answer
-
Filament Current
-
Anode Voltage
-
Focal Spot Size
-
Exposure Time
Question 10
Question
A function of the cathode is to:
Question 11
Question
The function of the rotor & the stator assembly is to:
Question 12
Question
Each of the following is a part of the cathode assembly except the:
Answer
-
Stator
-
Filament
-
Focusing Cup
-
Associated Wiring
Question 13
Question
Each of the following is a material of choice for the filament except:
Answer
-
Copper
-
Tungsten
-
Rhenium
-
Molybdenum
Question 14
Question
Filament materials are chosen for their:
1)High Melting Point
2)Low Melting Point
3)Ability to Vaporize Easily
4)High Atomic Number
Answer
-
Numbers 1 & 3 only
-
Numbers 2 & 3 only
-
Numbers 1 & 4 only
-
Numbers 2 & 4 only
-
Numbers 1, 3, & 4 only
Question 15
Question
The focusing cup has a _______ charge.
Question 16
Question
The purpose of the focusing cup is to:
Answer
-
Alter the filament size
-
Group the electrons for their passage to the anode
-
Regulate anode rotation speed
-
Increase the heat capacity
Question 17
Question
X-ray production occurs at the:
Answer
-
Anode
-
Cathode
-
Filament
-
Stator
Question 18
Question
To extend the x-ray tube life, the technologist should:
1)Perform warm up exposures at the beginning of each day
2)Extend the time that the boost current (prep) button is held down
3)Avoid repeated exposures at or near the tube's capacity
Answer
-
1 & 2 Only
-
1 & 3 Only
-
2 & 3 Only
-
1, 2, & 3
Question 19
Question
When the anode angle is less than 45 degrees:
Answer
-
Effective focal spot is smaller than the actual focal spot
-
The actual focal spot of smaller than the effective focal spot
-
The angle has no effect on focal spot
Question 20
Question
The exact point in the x-ray unit where x-ray photons are created in:
Answer
-
Focusing Cup
-
Rotor
-
Filament
-
Focal Spot
Question 21
Question
To cover a 14x17 inch field at 40 inch SID, a ________ target angle is required.
Answer
-
3 Degrees
-
5 Degrees
-
7 Degrees
-
12 Degrees
Question 22
Question
The use of line focus principle causes a problem known as the ________.
Answer
-
Anode Heel Effect
-
Space Charge Effect
-
Quantum Mottle
-
Vacuum
Question 23
Question
The disadvantage of a small target angle is:
Answer
-
Increased heat distribution & capacity
-
Greater field coverage
-
A greater anode heel effect
-
A more uniform radiographic density
Question 24
Question
Warming the x-ray tube:
1)Is never necessary
2)Prevents cracking of the anode
3)Helps maintain the vacuum within the tube
Question 25
Question
Pitting of the anode commonly occurs:
Answer
-
Consistent low but very long exposures
-
Consistent single, high exposures
-
Failure to warm up the tube
-
All the Above
Question 26
Question
Anode melt might occur in a specific area on the anode due to:
Question 27
Question
The anode heel effect:
Answer
-
Exists because some x-rays are absorbed by the surface of the anode
-
Is caused by the charge difference between the anode & cathode
-
Depends on the mA & kVp
-
Is reduced by dual focal spot
Question 28
Question
The line focus principle:
Question 29
Question
To best apply the anode heel effect, the anode should be placed towards which end of the body when performing an AP projection?
Answer
-
Towards the Knee
-
Towards the Hip
Question 30
Question
The vacuum created within the x-ray tube increase the efficiency of the tube's operation.
Question 31
Question
The cathode is the positive side of the x-ray tube.
Question 32
Question
The faster the anode rotation, the better the heat dissipation.
Question 33
Question
Many x-ray tubes have two filaments:
Answer
-
Because the second filament can be used as a spare part when the first is burned out
-
Provide two focal spots
-
To allow cooling of the filament by alternating exposures
-
To improve tube cooling by sharing the hear between two filaments
Question 34
Question
Modern general purpose radiographic equipment uses a stationary anode.
Question 35
Question
The most common diagnostic radiography target angle is 12 degrees.
Question 36
Question
The anode disk is composed of several different materials.
Question 37
Question
Changing the angle of the anode will change the effective focal spot.
Question 38
Question
One way to extend tube life is by using high mA stations when possible.