ST Monitoring Quiz

Descripción

ST monitoring via Philips monitor
Natalie Tredway
Test por Natalie Tredway, actualizado hace más de 1 año
Natalie Tredway
Creado por Natalie Tredway hace más de 9 años
278
1

Resumen del Recurso

Pregunta 1

Pregunta
ST analysis can be performed on the patient with 1:1 ventricular pacing.
Respuesta
  • True
  • False

Pregunta 2

Pregunta
The ST segment is:
Respuesta
  • Normally elevated in all leads
  • Normally isoelectric in all leads
  • Normally depressed in all leads
  • Damaged and ischemia causes the cells of the myocardium to become more or less excitable.

Pregunta 3

Pregunta
ST segment measurements are displayed and trended in:
Respuesta
  • mmHg
  • mm
  • kpal

Pregunta 4

Pregunta
The ST Leads that are being monitored depend on: (choose all that apply)
Respuesta
  • The leads available based on the lead set used and electrodes applied.
  • The Lead Placement selected - Standard or EASI
  • The number of leads displayed on the Main Screen.
  • The leads setup for monitoring.

Pregunta 5

Pregunta
When ST Analysis is turned off:
Respuesta
  • No ST analysis or arrhythmia analysis is performed.
  • No ST analysis is performed but arrhythmia analysis continues.

Pregunta 6

Pregunta
To turn off individual leads of ST:
Respuesta
  • Change to another screen
  • In the ST Analysis menu, select Setup ST Leads.
  • In the ECG menu, select New Lead Setup.
  • In the ST Analysis menu, select, select ST Analysis.

Pregunta 7

Pregunta
Selecting Alarms in the ST Analysis menu:
Respuesta
  • Turns on/off all ST alarms.
  • Allows you turn off individual ST alarms.
  • Silences ST alarms for 5 minutes.
  • Allows you to select single-lead or multi-lead alarms.

Pregunta 8

Pregunta
Multi-lead alarms will cause an alarm only when:
Respuesta
  • All monitored ST leads exceed the limits set for 15 seconds.
  • Three or more leads exceed the limits set for 15 seconds.
  • Two or more leads exceed the limits set for 1 minute.
  • Three or more leads exceed the limits set for 1 minute.

Pregunta 9

Pregunta
ST alarms set using Auto Limits Wide: (choose all that apply)
Respuesta
  • Changes the alarm limits based on the last 5 minutes of data.
  • Change as the patient’s condition changes
  • Changes the alarm limits based on the last 1 minute of data.
  • Will not change until you change them manually or use AutoLimits again.
  • Will not set alarm limits in pathological range.
Mostrar resumen completo Ocultar resumen completo

Similar

Tabla Periódica de los Elementos
maya velasquez
Phrasal verbs
John Goalkeeper
Fracciones y porcentajes
Esperanza Gesteira
FUNDAMENTOS DE LA LOGÍSTICA
amorsitoliss
NUMEROS RACIONALES - FRACCIONES
Jose Daniel Pauc
Mitosis y meiosis
wilmar252
Organizadores graficos
obvelasquezl34
AMBIENTES VIRTUALES DE APRENDIZAJE (AVA)
naansara1993
Tema 4, Los paisajes de España
Mercedes Graves
Uso De Los Tiempos Verbales
Laura -
Medicina Forense
Alfonso Tester