Pregunta 1
Pregunta
The nuclear force which holds electrons in “orbit” around an atom is called:
Respuesta
-
gravity
-
charge
-
valence force
-
binding energy
Pregunta 2
Pregunta
change from 2 mm Al filtration to 4 mm Al filtration
The curve shifts:
Respuesta
-
right
-
left
-
remains the same
Pregunta 3
Pregunta
Scatter radiation is mostly a result of the Photoelectric interaction.
Pregunta 4
Pregunta
change from 2 mm Al filtration to 4 mm Al filtration
The discrete (characteristic) curve shifts:
Respuesta
-
right
-
left
-
remains the same
Pregunta 5
Pregunta
change from high atomic number target material to low atomic number
The amplitude (height) of the emission spectrum curve would:
Respuesta
-
increase
-
decrease
-
remain the same
Pregunta 6
Pregunta
The Compton interaction is mainly responsible for radiographer dose.
Pregunta 7
Pregunta
change from 82 kVp to 72 kVp
The amplitude (height) of the emission spectrum curve would:
Respuesta
-
decrease
-
increase
-
remain the same
Pregunta 8
Pregunta
As the atomic number of an element increases, the energy required to remove an inner shell electron _____________________.
Respuesta
-
increases
-
decreases
-
is unchanged
Pregunta 9
Pregunta
change from high atomic number target material to low atomic number
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Pregunta 10
Pregunta
The end product(s) of the photoelectric effect is/are:
Respuesta
-
one ejected electron and an orbital vacancy
-
one ejected electron only
-
one scattered photon, one ejected electron, and an orbital vacancy
-
one scattered photon only
-
one orbital vacancy only
Pregunta 11
Pregunta
If an incident photon has an energy of 90 keV and undergoes a Compton interaction giving 20 keV to the ejected electron which had a binding energy of 2 keV, determine the energy of the scattered photon.
the energy is [blank_start]68[blank_end] keV
Pregunta 12
Pregunta
Which of the following is NOT an interaction between x-ray photons and matter?
Pregunta 13
Pregunta
The greatest amount of scatter is produced from the _____________________ interaction.
Respuesta
-
coherent
-
compton
-
characteristic
-
brems
Pregunta 14
Pregunta
Electromagnetic waves travel at the speed of light.
Pregunta 15
Pregunta
A high energy x-ray photon has a long wavelength.
Pregunta 16
Pregunta
An 80-kV photon undergoes a coherent scattering interaction with an orbital electron having a binding energy of 8 kV. What is the energy of the resulting scattered photon?
Pregunta 17
Pregunta
Filtration is the process of eliminating undesirable ____ by the insertion of absorbing materials into the primary beam.
Respuesta
-
low energy x-ray photons
-
high energy incident electrons
-
high energy x-ray photons
-
low energy incident electrons
Pregunta 18
Pregunta
label the parts of the x-ray beam spectrum graph
Pregunta 19
Pregunta
change from high atomic number target material to low atomic number
The curve shifts:
Respuesta
-
left
-
right
-
remains the same
Pregunta 20
Pregunta
If the number of protons within an atom’s nucleus is changed, which of the following must absolutely change:
Respuesta
-
number of electrons
-
element's name
-
electrical charge
-
number of neutrons
Pregunta 21
Pregunta
change from 82 kVp to 72 kVp
The curve shifts:
Respuesta
-
left
-
right
-
remain the same
Pregunta 22
Pregunta
Characteristic radiation forms a discrete emission spectrum.
Pregunta 23
Pregunta
The partial absorption of the x-ray beam as it passes through the body is best referred as:
Respuesta
-
absorption
-
brems
-
heterogeneity
-
attenuation
Pregunta 24
Pregunta
this interaction is a ____________ interaction
Pregunta 25
Pregunta
change from 82 kVp to 72 kVp
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Pregunta 26
Pregunta
A low energy brems x-ray results when the incident electron is strongly influenced by the nucleus.
Pregunta 27
Pregunta
change from high atomic number target material to low atomic number
The discrete (characteristic) curve shifts:
Respuesta
-
left
-
right
-
remains the same
Pregunta 28
Pregunta
change from 2 mm Al filtration to 4 mm Al filtration.
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Pregunta 29
Pregunta
If an incident x-ray photon ejects a k-shell electron with a binding energy of 37 keV, and the initial energy of the photon was 40 keV, the ejected electron leaves the atom with _____ keV of kinetic energy.
Pregunta 30
Pregunta
A photon of x-ray has an electrically positive charge.
Pregunta 31
Pregunta
Attenuation causes a gradual decrease in the number of primary x-rays that exit the patient.
Pregunta 32
Pregunta
Brems is produced by:
Respuesta
-
slowing down of an electron by the nucleus
-
ionization from ejection of outer shell electron
-
excitation and restabilization of an outer shell electron
-
creation of a photoelectron
Pregunta 33
Pregunta
change from 82 kVp to 72 kVp
The discrete (characteristic) curve shifts:
Respuesta
-
left
-
right
-
remain the same
Pregunta 34
Pregunta
In the diagnostic range, the majority of primary x-rays produced are:
Respuesta
-
brems
-
compton
-
photoelectric
-
characteristic
Pregunta 35
Pregunta
Radiation that turns 180 degrees in direction after interacting with matter is called:
Respuesta
-
coherent
-
backscatter
-
brems
-
characteristic
Pregunta 36
Pregunta
change from 2 mm Al filtration to 4 mm Al filtration
The amplitude (height) of the emission spectrum curve would:
Respuesta
-
increase
-
decrease
-
remain the same
Pregunta 37
Pregunta
If a projectile electron is traveling at 70 keV and is deviated by the nucleus with a resulting electron energy of 10 keV, determine the energy of emitted radiation = [blank_start]60[blank_end] keV
This is an example of what type of radiation? [blank_start]brems[blank_end] interaction
Pregunta 38
Pregunta
We used an original mAs of 32 and kVp of 80, list two separate techniques (one kVp & mAs then another kVp & mAs) we could use to double our exposure. (0.5 pt for each)
15% rule: [blank_start]92[blank_end] kVp @ [blank_start]32[blank_end] mAs
Doubling mAs: [blank_start]64[blank_end] mAs @ [blank_start]80[blank_end] kVp
Pregunta 39
Pregunta
label the x-ray tube
Respuesta
-
stator
-
rotor
-
glass envelope
-
filament
-
focusing cup
-
target
-
window
-
disk
Pregunta 40
Pregunta
Distortion can be a misrepresentation of
Respuesta
-
shape only
-
size only
-
detail
-
shape and size
Pregunta 41
Pregunta
Calculate the magnification factor when the SID is 40 in. and the OID is 5 in. round to 2 decimal places
[blank_start]1.14[blank_end]x
Pregunta 42
Pregunta
The difference between two adjacent shades of gray is:
Respuesta
-
distortion
-
subject contrast
-
spatial resolution
-
quantum mottle
Pregunta 43
Pregunta
Spatial resolution is improved when:
Respuesta
-
SID decreases
-
OID decreases
-
SOD decreases
-
OID increases
Pregunta 44
Pregunta
Foreshortening is present if:
Respuesta
-
The image of the object is shorter than the object
-
The image of the object is larger than the object
-
the tube and IR are in alignment
-
IR exposure is excessive
Pregunta 45
Pregunta
If 80 kVp, 400 mA, and 0.025 sec. produce a satisfactory radiograph, what mAs was used?
[blank_start]10[blank_end] mAs
Pregunta 46
Pregunta
The [blank_start]___[blank_end] the SID, the [blank_start]___[blank_end] the magnification.
Respuesta
-
lesser
-
greater
-
n/a
-
smaller
-
bigger
-
n/a
Pregunta 47
Pregunta
change from 20 mAs to 40 mAs
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Pregunta 48
Pregunta
As tube angle increases, shape distortion will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 49
Pregunta
What can negatively impact spatial resolution?
Respuesta
-
patient motion
-
small focal spot
-
minimal OID
-
small detectors
Pregunta 50
Pregunta
change from 3 mm Al filtration to 5 mm Al filtration
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Pregunta 51
Pregunta
Factors affecting spatial resolution include:
Respuesta
-
focal spot size
-
contrast scale
-
kVp
-
mAs
Pregunta 52
Pregunta
A good method for controlling motion when the patient is unable to cooperate is to:
Pregunta 53
Pregunta
The effect of mAs upon IR exposure is:
Respuesta
-
directly proportional
-
inversely proportional
-
dependent on kVp
Pregunta 54
Pregunta
If a radiograph using 40 mAs (400 mA at 0.10 sec.) produced a radiograph with satisfactory exposure, what new time should be used if mA is changed to 200?
[blank_start]0.2[blank_end] sec
Pregunta 55
Pregunta
Poor spatial resolution is caused by an unacceptable amount of
Respuesta
-
umbra
-
penumbra
-
IR Exposure
-
contrast
Pregunta 56
Pregunta
the unsharpness in an image is affected by all of the following except:
Respuesta
-
focal spot size
-
mAs
-
SOD
-
OID
Pregunta 57
Pregunta
If a projected image measures 3 in. and the magnification factor is 1.1, what is the size of the actual object? (you must label your answer with abbreviation)
[blank_start]2.7 in[blank_end]
Pregunta 58
Pregunta
The principal controlling technical factor of remnant beam contrast is
Respuesta
-
mAs
-
kVp
-
focal spot size
-
anode angle
Pregunta 59
Pregunta
As kVp increases, remnant beam contrast scale will _________________.
Respuesta
-
lengthen
-
shorten
-
remain unchanged
Pregunta 60
Pregunta
An increase in x-ray beam filtration
Respuesta
-
increases magnification
-
increases the average photon energy of the beam
-
decreases patient tissue density
-
increases the quantity of photons in the beam
Pregunta 61
Pregunta
A radiograph with few shades of gray exhibits
Respuesta
-
high contrast
-
medium contrast
-
low contrast
Pregunta 62
Pregunta
Calculate the magnification factor when the SID is 40 in. and the SOD is 25 in.
[blank_start]1.6[blank_end]x
Pregunta 63
Pregunta
As kVp decreases, spatial resolution will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 64
Pregunta
As filtration increases, IR exposure will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 65
Pregunta
If the following three exposures were obtained on the same patient (with no computer processing and/or on film) and everything is unchanged except for mAs, in which image would you expect mAs to have been the largest?
Pregunta 66
Pregunta
The principal controlling factor of radiographic image receptor exposure is:
Respuesta
-
mAs
-
kVp
-
focal spot size
-
tube angulation
Pregunta 67
Pregunta
With an SID of 40 inches, and an OID of 10 inches, what is the factor of magnification?
Pregunta 68
Pregunta
If 70 kVp, 100 mA, and 0.25 sec. produce a satisfactory radiograph, what mAs was used?
[blank_start]25[blank_end] mAs
Pregunta 69
Pregunta
If 10 mAs at 120 kVp is used for a radiograph and the remnant beam's scale of contrast is too long, determine the new technique to be used obtaining the same IR exposure.
new kVp: [blank_start]102[blank_end]
new mAs: [blank_start]20[blank_end]
Pregunta 70
Pregunta
Electronic "snow" on a display monitor or TV monitor image is a form of:
Respuesta
-
noise
-
scatter
-
distortion
-
IR exposure
Pregunta 71
Pregunta
As mAs increases, remnant beam contrast scale will _________________.
Respuesta
-
lengthen
-
shorten
-
remain unchanged
Pregunta 72
Pregunta
A change in kVp, as an exposure factor, affects
Pregunta 73
Pregunta
As kVp increases, the percentage of Compton interactions (relative to the total number of interactions) [blank_start]____[blank_end] and remnant beam contrast scale [blank_start]____[blank_end].
Respuesta
-
increase
-
decrease
-
N/A
-
lengthen
-
shorten
-
N/A
Pregunta 74
Pregunta
Size distortion is controlled by:
Respuesta
-
SID
-
OID
-
mAs
-
kVp
-
patient motion
Pregunta 75
Pregunta
Calculate the magnification factor when the SID is 72 in. and the SOD is 25 in. round to 2 places.
[blank_start]2.88[blank_end] x
Pregunta 76
Pregunta
As mAs increases, IR exposure will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 77
Pregunta
As mAs increases, shape distortion will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 78
Pregunta
As SID increases, spatial resolution will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 79
Pregunta
As focal spot size increases, spatial resolution will _______________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 80
Pregunta
The brightness of the image on a viewing monitor is the result of
Pregunta 81
Pregunta
What should you expect to occur with a decrease in kVp?
Pregunta 82
Pregunta
Insufficient mAs is likely to contribute to ___ in the image.
Respuesta
-
quantum mottle
-
motion
-
high spatial resolution
-
excessive OID
Pregunta 83
Pregunta
Using the 15% rule changes
Respuesta
-
distortion
-
IR exposure
-
remnant beam contrast
Pregunta 84
Pregunta
A radiograph of the forearm is produced using 4 mAs at 65 kVp. What kVp would be required to double the exposure? round up to whole number
[blank_start]75[blank_end] kVp
Pregunta 85
Pregunta
As OID increases, size distortion will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 86
Pregunta
If the signal is 16 and the noise is 4 on a hypothetical radiographic image, what is the Signal to Noise ratio?
Pregunta 87
Pregunta
Which of the following is equivalent to low penumbra?
Pregunta 88
Pregunta
Penumbra decreases as
Respuesta
-
SID increases
-
OID decreases
-
focal spot decreases
-
all of the above
Pregunta 89
Pregunta
If the mA is set at 300 and the total mAs produced is 15 mAs, the exposure time must be:
Respuesta
-
0.5 sec
-
0.05 sec
-
0.2 sec
-
2 sec
Pregunta 90
Pregunta
Size distortion in radiography can be ___ only.
Respuesta
-
magnification
-
foreshortening
-
elongation
-
minimization
Pregunta 91
Pregunta
What factor influences the average energy of photons reaching the IR?
Respuesta
-
mAs
-
kVp
-
mA
-
focal spot size
Pregunta 92
Pregunta
If 20 mAs at 100 kVp is used and the IR exposure needs to be doubled without altering the remnant beam contrast scale, determine the new technique that should be used.
new kVp: [blank_start]100[blank_end]
new mAs: [blank_start]40[blank_end]
Pregunta 93
Pregunta
which of the following are causes of scatter radiation
Respuesta
-
high kVp
-
low kVp
-
large part thickness
-
part density
-
large field sizes
-
small field sizes
-
large mAs
-
small mAs
Pregunta 94
Pregunta
what is the purpose of a grid
Respuesta
-
to absorb low energy photons before they reach the IR
-
to absorb scatter before it reaches IR
-
to reflect scatter away from IR
-
to absorb primary radiation before it reaches IR
Pregunta 95
Pregunta
match the type of grid with its definition
[blank_start]Stationary[blank_end] grid: attaches to cassette
[blank_start]Reciprocating[blank_end] grid: moves grid back and forth
[blank_start]Oscillating[blank_end] grid: move grid plate in circular motion
The [blank_start]reciprocating[blank_end] and [blank_start]oscillating[blank_end] grids make up the Potter-Bucky Diaphragm aka the [blank_start]bucky[blank_end]
Respuesta
-
Stationary
-
Reciprocating
-
Oscillating
-
reciprocating
-
oscillating
-
bucky
Pregunta 96
Pregunta
the [blank_start]higher[blank_end] the grid ratio, the more effective the grid is at removing scatter
Pregunta 97
Pregunta
We used 10 mAs at 78 kV on a KUB. Convert this non-grid technique to a 10:1 grid
[blank_start]50[blank_end] mAs @ [blank_start]78[blank_end] kV
Pregunta 98
Pregunta
Convert 60 mAs at 92 kV using a 8:1 grid to an 5:1 grid.
[blank_start]30[blank_end] mAs @ [blank_start]92[blank_end] kV
Pregunta 99
Pregunta
increased lead content = [blank_start]decreased[blank_end] scatter = contrast scale [blank_start]shortens[blank_end] = image contrast [blank_start]increases[blank_end]
Respuesta
-
decreased
-
increased
-
same
-
shortens
-
lengthens
-
remains the same
-
increases
-
decreases
-
remains the same
Pregunta 100
Pregunta
grid cut off is due to the absorption of
Respuesta
-
scatter radiation
-
remnant radiation
-
primary radiation
-
secondary radiation
Pregunta 101
Pregunta
label the type of grid
Respuesta
-
short axis
-
linear
-
cross-hatch
Pregunta 102
Pregunta
label the type of grid
Pregunta 103
Pregunta
What is the grid radius based on?
Respuesta
-
SID
-
OID
-
Tube angle
-
Object thickness
Pregunta 104
Pregunta
fill out the table and GCF equation, include x in multiplier.
Respuesta
-
No grid
-
x1
-
5:1
-
x2
-
6:1
-
x3
-
8:1
-
x4
-
10:1 or 12:1
-
x5
-
16:1
-
x6
-
mAs2
-
mAs1
-
GCF2
-
GCF1
Pregunta 105
Pregunta
What causes the moire effect?
Respuesta
-
the grid lines are parallel to scan lines of processor
-
the grid lines are perp to scan lines of processor
-
misalignment of the grid and beam
-
incorrect SID used
Pregunta 106
Pregunta
This image is an example of
Pregunta 107
Pregunta
Match the image with its grid error
Pregunta 108
Pregunta
The air gap technique is when OID is [blank_start]increases[blank_end], so scatter on the IR [blank_start]decreases[blank_end]
Respuesta
-
increased
-
decreased
-
remains the same
-
increases
-
decreases
-
remains the same
Pregunta 109
Pregunta
When generator power and rectification increase
Respuesta
-
quantity of current increases
-
quantity of current decreases
-
quality of current increases
-
quality of current decreases
-
current is not affected
Pregunta 110
Pregunta
high frequency generators [blank_start]reduce[blank_end] the ripple of the wave form
with each increase in generator power, the exposure rate of the x-ray beam [blank_start]increases[blank_end]
These changes [blank_start]increase[blank_end] the effective mA, which allows for a [blank_start]reduction[blank_end] in technique
High power generators also [blank_start]increase[blank_end] the average kV
Generally, the set mAs could be [blank_start]cut in half[blank_end] when changing from a single phase to a 3 phase/HF
Respuesta
-
reduce
-
increase
-
don't change
-
increases
-
decreases
-
remains the same
-
increase
-
decrease
-
do not change
-
decrease
-
increase
-
no change
-
increase
-
decrease
-
do not change
-
cut in half
-
doubled
-
unchanged
Pregunta 111
Pregunta
generator power [blank_start]lengthens[blank_end] remnant contrast scale
generator power [blank_start]does not affect[blank_end] spatial resolution
generator power [blank_start]does not affect[blank_end] size distortion
generator power [blank_start]does not affect[blank_end] shape distortion
Respuesta
-
does not affect
-
shortens
-
lengthens
-
does not affect
-
increases
-
decreases
-
does not affect
-
increases
-
decreases
-
does not affect
-
increases
-
decreases
Pregunta 112
Pregunta
which of the following are considered inherent filtration
Pregunta 113
Pregunta
which of the following are considered added filtration
Pregunta 114
Pregunta
The minimum total filtration equivalent is [blank_start]2.5[blank_end] mm Al
- [blank_start]1.5[blank_end] mm of inherent filtration
- [blank_start]1[blank_end] mm added filtration
Pregunta 115
Pregunta
What is the unit for directly measuring x-ray beam penetration
Respuesta
-
Half-Value Layer
-
Half-Vacancy Layer
-
mm Al
-
mm Hg
-
mm Pb
Pregunta 116
Pregunta
A Potter-Bucky diaphragm does not move
Pregunta 117
Pregunta
Higher frequency grids have [blank_start]thinner[blank_end] lead strips
Respuesta
-
thinner
-
thicker
-
no difference in
Pregunta 118
Pregunta
What is the most critical factor in a grid's efficiency?
Respuesta
-
grid frequency
-
grid ratio
-
lead content of grids
-
grid dimension
Pregunta 119
Pregunta
Grid radius is the distance between the grid and the [blank_start]convergence line[blank_end].
Pregunta 120
Pregunta
Match the term with its definition
[blank_start]Grid ratio[blank_end] : the proportion of the height to the width of the interspaced between the lead strips
[blank_start]Lead Content of Grid[blank_end] : mass per unit area, as this increases so does the removal of scatter
[blank_start]Grid Radius[blank_end]: the distance between the grid and the convergent line or point aka the correct SID
[blank_start]Grid Frequency[blank_end]: the number of lead strips counted per inch/cm
Respuesta
-
Grid ratio
-
Lead Content of Grid
-
Grid Radius
-
Grid Frequency
Pregunta 121
Pregunta
What is the purpose of a compensating filter?
Pregunta 122
Pregunta
Which of the follower are the purposes for field size limitation
Respuesta
-
minimize radiation exposure to the tech
-
minimize radiation exposure to the patient
-
change subject contrast in the remnant beam
-
preserve subject contrast in the remnant beam
-
preserve image sharpness
Pregunta 123
Pregunta
increased collimation (smaller size) = [blank_start]decreased[blank_end] scatter = [blank_start]shortened[blank_end] contrast scale
increased collimation = [blank_start]decreased[blank_end] IR exposure
Respuesta
-
increased
-
decreased
-
same amount of
-
lengthened
-
shortened
-
same
-
decreased
-
increased
-
same
Pregunta 124
Pregunta
match the types of field size limitations
Pregunta 125
Pregunta
The purpose of PBL is
Respuesta
-
to allow techs to not worry about collimation
-
ensure the field size is never larger than cassette
-
allow for easy collimation
-
automatically set field size
Pregunta 126
Pregunta 127
Pregunta
Positive Contrast Agents
-[blank_start]Radiopaque[blank_end] ([blank_start]white[blank_end])
-Based upon [blank_start]Barium or Iodine[blank_end]
Negative Contrast Agents
-[blank_start]Radiolucent[blank_end] ([blank_start]black[blank_end])
-Are all [blank_start]gasses[blank_end]
Respuesta
-
Radiopaque
-
Radiolucent
-
white
-
black
-
Barium or Iodine
-
Barium or Calcium
-
Lead or Tungsten
-
Tungsten or Molybdenum
-
Calcium or Iodine
-
Barium or Tungsten
-
Radiopaque
-
Radiolucent
-
white
-
black
-
gasses
-
water
-
tissues
-
bones
-
barium
-
iodine
Pregunta 128
Pregunta
Additive diseases = [blank_start]increased radiopacity of tissue[blank_end] = [blank_start]increased[blank_end] technique
Destructive diseases = [blank_start]increased radiolucency of tissue[blank_end] = [blank_start]decreased[blank_end] technique
Respuesta
-
increased radiopacity of tissue
-
increased radiolucency of tissue
-
decreased radiopacity of tissue
-
decreased radiolucency of tissue
-
increased
-
decreased
-
same
-
increased radiopacity of tissue
-
increased radiolucency of tissue
-
decreased radiopacity of tissue
-
decreased radiolucency of tissue
-
increased
-
decreased
-
same
Pregunta 129
Pregunta
which of the following are additive diseases
Respuesta
-
Emphysema
-
Pleural Effusion
-
Edema
-
Osteoporosis
-
Pneumothorax
-
Pneumonia
-
Cardiomegaly
-
Cancer
-
increase fluid, bone or metal content
-
increase gas or fat, decrease normal fluid or bone
Pregunta 130
Pregunta
which of the following are destructive diseases
Respuesta
-
Emphysema
-
Pleural Effusion
-
Edema
-
Osteoporosis
-
Pneumothorax
-
Pneumonia
-
Cardiomegaly
-
Cancer
-
increase fluid, bone or metal content
-
increase gas or fat, decrease normal fluid or bone
Pregunta 131
Pregunta
For soft tissue techniques, kVp is
Pregunta 132
Pregunta
For patients with a fiberglass cast/splint, the technique may be altered by
Respuesta
-
an increase in mAs by 50%
-
an increase in mAs by 20%
-
an increase in mAs by 8%
-
an increase in kVp by 50%
-
an increase in kVp by 20%
-
an increase in kVp by 8%
Pregunta 133
Pregunta
the longer the SID, the [blank_start]sharper[blank_end] the detail
the longer the SID, the [blank_start]less[blank_end] magnification
the longer the SID, the [blank_start]no effect on[blank_end] shape distortion
Respuesta
-
sharper
-
fuzzier
-
no effect on
-
more
-
less
-
no effect on
-
more
-
less
-
no effect on
Pregunta 134
Pregunta
fill in the formula for the inverse square law
Pregunta 135
Pregunta
fill in the formula for the exposure maintenance law
Pregunta 136
Pregunta
Increased OID = [blank_start]Decreased[blank_end] Scatter = [blank_start]shorter[blank_end] contrast scale
Increased OID = [blank_start]decreased[blank_end] scatter = [blank_start]decreased[blank_end] IR exposure
Increased OID = [blank_start]Decreased[blank_end] sharpness
Increased OID = [blank_start]Increased[blank_end] magnification
Increased OID = [blank_start]No effect on[blank_end] shape distortion
Respuesta
-
decreased
-
increased
-
no effect on
-
shorter
-
longer
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
Pregunta 137
Pregunta
If 40 mAs at 76 kVp is used and the amount of exposure needs to be cut in half, determine the new technique that should be used if you want to use the 15% rule. round to 1 decimal place if necessary
[blank_start]64.6[blank_end] kVp @ [blank_start]40[blank_end] mAs
Pregunta 138
Pregunta
If 80mAs at 70kVp is used and the IR exposure needs to be doubled and the remnant beam scale of contrast lengthened, determine the new technique that should be used. round to one decimal if necessary
new kVp: [blank_start]80.5[blank_end]
new mAs: [blank_start]80[blank_end]
Pregunta 139
Pregunta
A radiograph of the forearm is produced using 4 mAs at 65 kVp. round to 1 decimal place if necessary
new kVp to double IR exposure: [blank_start]74.8[blank_end]
new kVp to halve IR exposure: [blank_start]55.3[blank_end]
Pregunta 140
Pregunta
Volume of the irradiated tissue increases as
Pregunta 141
Pregunta
The half-value layer (HVL) is used to measure
Pregunta 142
Pregunta
Photon absorption within the body occurs as the result of
Pregunta 143
Pregunta
The [blank_start]____[blank_end] the photon energy, the [blank_start]____[blank_end] the percentage of interactions that will be Compton interactions.
Respuesta
-
higher
-
lower
-
higher
-
lower
-
unaffected
Pregunta 144
Pregunta
If OID increases, the remnant beam contrast scale [blank_start]_________________[blank_end].
Respuesta
-
lengthens
-
shortens
-
is unaffected
Pregunta 145
Pregunta
The computer distinguishes the anatomy of interest from background exposure outside collimated field during:
Pregunta 146
Pregunta
How dark or light a digital image appears on a display monitor is known as
Respuesta
-
contrast resolution
-
brightness
-
spatial resolution
-
density
Pregunta 147
Pregunta
In generic digital imaging terms, the average brightness level of the image is called the:
Respuesta
-
window level
-
window width
-
density
-
scale of contrast
Pregunta 148
Pregunta
In a digital image histogram graph, the vertical height of any data point along the plotted curve (measured against the y axis of the graph), indicates the:
Pregunta 149
Pregunta
A barium study projection will likely require the computer to locate both Smin and Smax.
Pregunta 150
Pregunta
On an image histogram, an unusual spike to the left of the main bell-shaped lobe of the curve most likely represents:
Respuesta
-
osteoporosis
-
background exposure
-
a large metallic object
-
air in the lungs
Pregunta 151
Pregunta
Dependent on the manufacturer’s algorithm, a value of interest (VOI) can accentuate a certain tissue thickness or bone.
Pregunta 152
Pregunta
The larger the number of pixels in an image matrix, the:
Respuesta
-
shorter the acquisition time
-
greater the resolution of the image
-
greater the number of gray shades in the image
-
all answers are correct
Pregunta 153
Pregunta
Which of the following is an example of a Field Uniformity Correction?
Respuesta
-
creation of a best fit line for the raw detector data
-
a flattening of the histogram curve as needed
-
application of a kernel across the matrix
-
evening out of brightness to account for irregularities in the hardware of the IT across the image field
Pregunta 154
Pregunta
Many imaging systems present the histogram with an averaged gray scale curve overlying the histogram. The steeper the slope:
Pregunta 155
Pregunta
Digital radiography images that are underexposed have a high SNR.
Pregunta 156
Pregunta
The development of which of the following enabled DR systems to use a flat-shaped receptor plate:
Pregunta 157
Pregunta
In digital imaging processing, what is segmentation?
Respuesta
-
a graph plotting signal intensity for each pixel
-
identification of the exposed area of the IR
-
the anatomy on which the exam is performed
-
a look up table
Pregunta 158
Pregunta
Generally speaking, the smaller the pixels, the higher the spatial resolution.
Pregunta 159
Pregunta
Which of the following terms, all contractions, most accurately describes the individual hardware elements of a DR flat panel receptor that detect radiation in the remnant x-ray beam?
Respuesta
-
the dexel
-
the pixel
-
the voxel
-
the rexel
Pregunta 160
Pregunta
The selection of a narrow window width will:
Respuesta
-
decrease the size of each pixel
-
decrease overall brightness of the image
-
increase contrast in the image
-
increase magnification of the central portion of the image
Pregunta 161
Pregunta
Which of the following modalities create images with the highest spatial resolution?
Respuesta
-
ct
-
nuclear medicine
-
ultrasound
-
digital radiography
Pregunta 162
Pregunta
Adjustment of the window width of the image will change
Respuesta
-
the brightness of the image
-
the level of IR exposure
-
the spatial resolution of the image
-
the scale of contrast of the image
Pregunta 163
Pregunta
The smaller the pixels, the lower the spatial resolution.
Pregunta 164
Pregunta
For digital imaging, each image displayed consists of a discrete number of rows and columns of picture elements called the:
Respuesta
-
signal plate
-
display table
-
matrix
-
detector array
Pregunta 165
Pregunta
In digital radiography, image contrast displayed on the monitor is controlled exclusively by kilovoltage.
Pregunta 166
Pregunta
The larger the number of pixels in an image matrix, the:
Respuesta
-
greater the number of gray shades in the image
-
shorter the acquisition time
-
greater the spatial resolution in the image
-
all answers are correct
-
no answers are correct
Pregunta 167
Pregunta
A system with a high S/N ratio displays:
Respuesta
-
low noise and high signal
-
low signal and high noise
-
equal amounts of signal and noise
-
no signal and no noise
Pregunta 168
Pregunta
Scatter radiation adds exposure to the image receptor.
Pregunta 169
Pregunta
Which factor influences the average energy of photons reaching the IR?
Respuesta
-
kVp
-
mA
-
time
-
focal spot size
Pregunta 170
Pregunta
As kVp decreases, spatial resolution will _________________.
Respuesta
-
increase
-
decrease
-
remain unchanged
Pregunta 171
Pregunta
In the diagnostic range, the majority of primary x-rays produced are:
Respuesta
-
characteristic
-
photoelectric
-
brems
-
compton
Pregunta 172
Pregunta
As kVp increases,
Pregunta 173
Pregunta
The amount of scatter reaching the image receptor is affected by all of the following EXCEPT:
Respuesta
-
SID
-
Amount of Collimation
-
kVp
-
grid ratio
Pregunta 174
Pregunta
The larger the number of pixels in an image matrix, the:
Pregunta 175
Pregunta
The higher the DQE, the higher the dose required to get an optimal image.
Pregunta 176
Pregunta
As the lead content of a grid increases,
Respuesta
-
image receptor exposure increases
-
the contrast decreases
-
The ability of the grid to remove scatter increases.
-
The ability of the grid to remove scatter decreases
Pregunta 177
Pregunta
In production of a radiograph, which of the following is considered the controlling factor of beam penetration and remnant beam scale of contrast?
Pregunta 178
Pregunta
The principal controlling factor of radiographic image receptor exposure is:
Respuesta
-
kVp
-
mAs
-
focal spot size
-
tube angulation
Pregunta 179
Pregunta
Image level parameters (window level) control image
Respuesta
-
brightness
-
contrast
-
density
-
darkness
Pregunta 180
Pregunta
All of the following are true of pixel pitch except:
Respuesta
-
it is measured from center-center of pixels
-
it is affected by changes in the matrix size
-
it is determined by the size of the dels in flat panel systems
-
it is not related to spatial resolution
Pregunta 181
Pregunta
Which of the following is NOT a component of a detector element?
Respuesta
-
capacitor
-
charged coupled device
-
semiconductor layer
-
thin film transistor
Pregunta 182
Pregunta
A phosphor screen or scintillator
Respuesta
-
fluoresces when exposed to light
-
fluoresces when exposed to xrays
-
ionizes selenium when exposed to x-rays
-
ionizes selenium when exposed to light
Pregunta 183
Pregunta
The percentage of the surface area of a flat panel detector element devoted to the semiconductor layer is:
Respuesta
-
DQE
-
Fill factor
-
detector ratio
-
capacitance
Pregunta 184
Pregunta
The purpose of the ADC is to convert the _______ signal to a _________ signal.
Respuesta
-
analog; digital
-
digital; analog
-
analog; detector
-
acquired; detected
-
detector; digital
Pregunta 185
Pregunta
Digital radiography images that are underexposed can show quantum mottle.
Pregunta 186
Pregunta
In a DR flat panel image receptor, which electronic devices directly account for the plate’s ability to store a latent image until it is sent into the computer?
Respuesta
-
capacitor
-
tfts
-
data lines
-
gate lines
-
semiconductor layer
Pregunta 187
Pregunta
An x-ray unit with a Modulation Transfer Function (MTF) of 0.2 is said to have perfect contrast resolution.
Pregunta 188
Pregunta
What is the size of each pixel for an image with a 1024x1024 matrix using a 30x30 cm field of view:
Pregunta 189
Pregunta
The ratio of SID to SOD is the primary control for:
Pregunta 190
Pregunta
In a DR flat panel image receptor, most of the square area of each del is dedicated to what portion of the detector element:
Respuesta
-
the semiconductor layer
-
the gate lines
-
the capacitors
-
the tfts
Pregunta 191
Pregunta
The ionization chamber AEC is located behind the IR on the wall Bucky.
Pregunta 192
Pregunta
As matrix size increases, the image's spatial resolution increases.
Pregunta 193
Pregunta
Direct conversion digital imaging systems include which of the following?
Respuesta
-
CCD
-
PSP
-
amorphous Se tft
-
amorphous Si tft
Pregunta 194
Pregunta
A CMOS imaging system would have
Respuesta
-
microlens optics
-
charged couple arrays
-
amorphous selenium
-
a helium red laser
Pregunta 195
Pregunta
Which of the following is NOT one of the three steps to digitization of an analog image?
Respuesta
-
Sampling
-
Quantization
-
Scanning
-
Initiation
Pregunta 196
Pregunta
The maximum range of pixel values a hardware device can store is:
Respuesta
-
voxel depth
-
bit depth
-
gray scale
-
gray level
Pregunta 197
Pregunta
A radiograph is made using the following factors:
100 mA 8:1 grid
1.0 Sec 8x10 collimated field size
70 kVp large focal spot
50" SID 3mm Al
2" OID suspended respiration
In each of the following, if only the proposed change is made, how will spatial resolution be affected?
- Use of a 10x12 field size: [blank_start]no change[blank_end]
- Use of no grid: [blank_start]no change[blank_end]
- Use 200 mA: [blank_start]no change[blank_end]
- Use 4" OID: [blank_start]decrease[blank_end]
- Use small focal spot: [blank_start]increase[blank_end]
- Use 4mm Al: [blank_start]no change[blank_end]
-Use 4 seconds: [blank_start]decrease[blank_end]
Respuesta
-
no change
-
increase
-
decrease
-
no change
-
increase
-
decrease
-
no change
-
increase
-
decrease
-
decrease
-
increase
-
no change
-
increase
-
decrease
-
no change
-
no change
-
increase
-
decrease
-
decrease
-
increase
-
no change
Pregunta 198
Pregunta
A radiograph is made using the following factors:
100 mA 8:1 grid
1.0 Sec 8x10 collimated field size
70 kVp large focal spot
50" SID 3mm Al
2" OID suspended respiration
In each of the following, if only the proposed change is made, how will IR exposure be affected?
- Use of a 10x12 field size: [blank_start]increased[blank_end]
- Use of no grid: [blank_start]increased[blank_end]
- Use 200 mA: [blank_start]increased[blank_end]
- Use 2.0 seconds: [blank_start]increased[blank_end]
- Use 4" OID: [blank_start]decreased[blank_end]
- Use 60" SID: [blank_start]decreased[blank_end]- Use 90 kVp: [blank_start]increased[blank_end]
Respuesta
-
increased
-
decreased
-
no change
-
increased
-
decreased
-
no change
-
increased
-
decreased
-
no change
-
increased
-
decreased
-
no change
-
decreased
-
increased
-
no change
-
decreased
-
increased
-
no change
-
increased
-
decreased
-
no change
Pregunta 199
Pregunta
A radiograph is made using the following factors:
500 mA
2.5 sec
130 kVp
36" SID
6" OID
2000x1000 matrix size
Window Width = 800
No grid
5x8 collimated field size
small focal spot
3mm Al
0.1mm pixel size
Window Level = 65
Changes are proposed below. In each of the following, if only the proposed change is made, how will spatial resolution be affected?
- Collimate to 10x10; [blank_start]n/a[blank_end]
- Use a grid; [blank_start]n/a[blank_end]
- Use 72" SID; [blank_start]increase[blank_end]
- Use 1" OID; [blank_start]increase[blank_end]
-Window level = 165; [blank_start]n/a[blank_end]
- Use 3600x3600 matrix size; [blank_start]increase[blank_end]
- Use 0.1cm pixel size; [blank_start]decrease[blank_end]
Respuesta
-
n/a
-
increase
-
decrease
-
n/a
-
increase
-
decrease
-
increase
-
decrease
-
n/a
-
increase
-
decrease
-
n/a
-
n/a
-
increase
-
decrease
-
increase
-
decrease
-
n/a
-
decrease
-
increase
-
n/a