Question 1
Question
The nuclear force which holds electrons in “orbit” around an atom is called:
Answer
-
gravity
-
charge
-
valence force
-
binding energy
Question 2
Question
change from 2 mm Al filtration to 4 mm Al filtration
The curve shifts:
Answer
-
right
-
left
-
remains the same
Question 3
Question
Scatter radiation is mostly a result of the Photoelectric interaction.
Question 4
Question
change from 2 mm Al filtration to 4 mm Al filtration
The discrete (characteristic) curve shifts:
Answer
-
right
-
left
-
remains the same
Question 5
Question
change from high atomic number target material to low atomic number
The amplitude (height) of the emission spectrum curve would:
Answer
-
increase
-
decrease
-
remain the same
Question 6
Question
The Compton interaction is mainly responsible for radiographer dose.
Question 7
Question
change from 82 kVp to 72 kVp
The amplitude (height) of the emission spectrum curve would:
Answer
-
decrease
-
increase
-
remain the same
Question 8
Question
As the atomic number of an element increases, the energy required to remove an inner shell electron _____________________.
Answer
-
increases
-
decreases
-
is unchanged
Question 9
Question
change from high atomic number target material to low atomic number
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Question 10
Question
The end product(s) of the photoelectric effect is/are:
Answer
-
one ejected electron and an orbital vacancy
-
one ejected electron only
-
one scattered photon, one ejected electron, and an orbital vacancy
-
one scattered photon only
-
one orbital vacancy only
Question 11
Question
If an incident photon has an energy of 90 keV and undergoes a Compton interaction giving 20 keV to the ejected electron which had a binding energy of 2 keV, determine the energy of the scattered photon.
the energy is [blank_start]68[blank_end] keV
Question 12
Question
Which of the following is NOT an interaction between x-ray photons and matter?
Question 13
Question
The greatest amount of scatter is produced from the _____________________ interaction.
Answer
-
coherent
-
compton
-
characteristic
-
brems
Question 14
Question
Electromagnetic waves travel at the speed of light.
Question 15
Question
A high energy x-ray photon has a long wavelength.
Question 16
Question
An 80-kV photon undergoes a coherent scattering interaction with an orbital electron having a binding energy of 8 kV. What is the energy of the resulting scattered photon?
Question 17
Question
Filtration is the process of eliminating undesirable ____ by the insertion of absorbing materials into the primary beam.
Answer
-
low energy x-ray photons
-
high energy incident electrons
-
high energy x-ray photons
-
low energy incident electrons
Question 18
Question
label the parts of the x-ray beam spectrum graph
Question 19
Question
change from high atomic number target material to low atomic number
The curve shifts:
Answer
-
left
-
right
-
remains the same
Question 20
Question
If the number of protons within an atom’s nucleus is changed, which of the following must absolutely change:
Answer
-
number of electrons
-
element's name
-
electrical charge
-
number of neutrons
Question 21
Question
change from 82 kVp to 72 kVp
The curve shifts:
Answer
-
left
-
right
-
remain the same
Question 22
Question
Characteristic radiation forms a discrete emission spectrum.
Question 23
Question
The partial absorption of the x-ray beam as it passes through the body is best referred as:
Answer
-
absorption
-
brems
-
heterogeneity
-
attenuation
Question 24
Question
this interaction is a ____________ interaction
Question 25
Question
change from 82 kVp to 72 kVp
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Question 26
Question
A low energy brems x-ray results when the incident electron is strongly influenced by the nucleus.
Question 27
Question
change from high atomic number target material to low atomic number
The discrete (characteristic) curve shifts:
Answer
-
left
-
right
-
remains the same
Question 28
Question
change from 2 mm Al filtration to 4 mm Al filtration.
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Question 29
Question
If an incident x-ray photon ejects a k-shell electron with a binding energy of 37 keV, and the initial energy of the photon was 40 keV, the ejected electron leaves the atom with _____ keV of kinetic energy.
Question 30
Question
A photon of x-ray has an electrically positive charge.
Question 31
Question
Attenuation causes a gradual decrease in the number of primary x-rays that exit the patient.
Question 32
Question
Brems is produced by:
Answer
-
slowing down of an electron by the nucleus
-
ionization from ejection of outer shell electron
-
excitation and restabilization of an outer shell electron
-
creation of a photoelectron
Question 33
Question
change from 82 kVp to 72 kVp
The discrete (characteristic) curve shifts:
Answer
-
left
-
right
-
remain the same
Question 34
Question
In the diagnostic range, the majority of primary x-rays produced are:
Answer
-
brems
-
compton
-
photoelectric
-
characteristic
Question 35
Question
Radiation that turns 180 degrees in direction after interacting with matter is called:
Answer
-
coherent
-
backscatter
-
brems
-
characteristic
Question 36
Question
change from 2 mm Al filtration to 4 mm Al filtration
The amplitude (height) of the emission spectrum curve would:
Answer
-
increase
-
decrease
-
remain the same
Question 37
Question
If a projectile electron is traveling at 70 keV and is deviated by the nucleus with a resulting electron energy of 10 keV, determine the energy of emitted radiation = [blank_start]60[blank_end] keV
This is an example of what type of radiation? [blank_start]brems[blank_end] interaction
Question 38
Question
We used an original mAs of 32 and kVp of 80, list two separate techniques (one kVp & mAs then another kVp & mAs) we could use to double our exposure. (0.5 pt for each)
15% rule: [blank_start]92[blank_end] kVp @ [blank_start]32[blank_end] mAs
Doubling mAs: [blank_start]64[blank_end] mAs @ [blank_start]80[blank_end] kVp
Question 39
Question
label the x-ray tube
Answer
-
stator
-
rotor
-
glass envelope
-
filament
-
focusing cup
-
target
-
window
-
disk
Question 40
Question
Distortion can be a misrepresentation of
Answer
-
shape only
-
size only
-
detail
-
shape and size
Question 41
Question
Calculate the magnification factor when the SID is 40 in. and the OID is 5 in. round to 2 decimal places
[blank_start]1.14[blank_end]x
Question 42
Question
The difference between two adjacent shades of gray is:
Answer
-
distortion
-
subject contrast
-
spatial resolution
-
quantum mottle
Question 43
Question
Spatial resolution is improved when:
Answer
-
SID decreases
-
OID decreases
-
SOD decreases
-
OID increases
Question 44
Question
Foreshortening is present if:
Answer
-
The image of the object is shorter than the object
-
The image of the object is larger than the object
-
the tube and IR are in alignment
-
IR exposure is excessive
Question 45
Question
If 80 kVp, 400 mA, and 0.025 sec. produce a satisfactory radiograph, what mAs was used?
[blank_start]10[blank_end] mAs
Question 46
Question
The [blank_start]___[blank_end] the SID, the [blank_start]___[blank_end] the magnification.
Answer
-
lesser
-
greater
-
n/a
-
smaller
-
bigger
-
n/a
Question 47
Question
change from 20 mAs to 40 mAs
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Question 48
Question
As tube angle increases, shape distortion will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 49
Question
What can negatively impact spatial resolution?
Answer
-
patient motion
-
small focal spot
-
minimal OID
-
small detectors
Question 50
Question
change from 3 mm Al filtration to 5 mm Al filtration
Which of the following best demonstrates the emission spectrum graph shift?
Red curve = new curve
Black curve = original curve
Question 51
Question
Factors affecting spatial resolution include:
Answer
-
focal spot size
-
contrast scale
-
kVp
-
mAs
Question 52
Question
A good method for controlling motion when the patient is unable to cooperate is to:
Question 53
Question
The effect of mAs upon IR exposure is:
Answer
-
directly proportional
-
inversely proportional
-
dependent on kVp
Question 54
Question
If a radiograph using 40 mAs (400 mA at 0.10 sec.) produced a radiograph with satisfactory exposure, what new time should be used if mA is changed to 200?
[blank_start]0.2[blank_end] sec
Question 55
Question
Poor spatial resolution is caused by an unacceptable amount of
Answer
-
umbra
-
penumbra
-
IR Exposure
-
contrast
Question 56
Question
the unsharpness in an image is affected by all of the following except:
Answer
-
focal spot size
-
mAs
-
SOD
-
OID
Question 57
Question
If a projected image measures 3 in. and the magnification factor is 1.1, what is the size of the actual object? (you must label your answer with abbreviation)
[blank_start]2.7 in[blank_end]
Question 58
Question
The principal controlling technical factor of remnant beam contrast is
Answer
-
mAs
-
kVp
-
focal spot size
-
anode angle
Question 59
Question
As kVp increases, remnant beam contrast scale will _________________.
Answer
-
lengthen
-
shorten
-
remain unchanged
Question 60
Question
An increase in x-ray beam filtration
Answer
-
increases magnification
-
increases the average photon energy of the beam
-
decreases patient tissue density
-
increases the quantity of photons in the beam
Question 61
Question
A radiograph with few shades of gray exhibits
Answer
-
high contrast
-
medium contrast
-
low contrast
Question 62
Question
Calculate the magnification factor when the SID is 40 in. and the SOD is 25 in.
[blank_start]1.6[blank_end]x
Question 63
Question
As kVp decreases, spatial resolution will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 64
Question
As filtration increases, IR exposure will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 65
Question
If the following three exposures were obtained on the same patient (with no computer processing and/or on film) and everything is unchanged except for mAs, in which image would you expect mAs to have been the largest?
Question 66
Question
The principal controlling factor of radiographic image receptor exposure is:
Answer
-
mAs
-
kVp
-
focal spot size
-
tube angulation
Question 67
Question
With an SID of 40 inches, and an OID of 10 inches, what is the factor of magnification?
Question 68
Question
If 70 kVp, 100 mA, and 0.25 sec. produce a satisfactory radiograph, what mAs was used?
[blank_start]25[blank_end] mAs
Question 69
Question
If 10 mAs at 120 kVp is used for a radiograph and the remnant beam's scale of contrast is too long, determine the new technique to be used obtaining the same IR exposure.
new kVp: [blank_start]102[blank_end]
new mAs: [blank_start]20[blank_end]
Question 70
Question
Electronic "snow" on a display monitor or TV monitor image is a form of:
Answer
-
noise
-
scatter
-
distortion
-
IR exposure
Question 71
Question
As mAs increases, remnant beam contrast scale will _________________.
Answer
-
lengthen
-
shorten
-
remain unchanged
Question 72
Question
A change in kVp, as an exposure factor, affects
Question 73
Question
As kVp increases, the percentage of Compton interactions (relative to the total number of interactions) [blank_start]____[blank_end] and remnant beam contrast scale [blank_start]____[blank_end].
Answer
-
increase
-
decrease
-
N/A
-
lengthen
-
shorten
-
N/A
Question 74
Question
Size distortion is controlled by:
Answer
-
SID
-
OID
-
mAs
-
kVp
-
patient motion
Question 75
Question
Calculate the magnification factor when the SID is 72 in. and the SOD is 25 in. round to 2 places.
[blank_start]2.88[blank_end] x
Question 76
Question
As mAs increases, IR exposure will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 77
Question
As mAs increases, shape distortion will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 78
Question
As SID increases, spatial resolution will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 79
Question
As focal spot size increases, spatial resolution will _______________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 80
Question
The brightness of the image on a viewing monitor is the result of
Question 81
Question
What should you expect to occur with a decrease in kVp?
Question 82
Question
Insufficient mAs is likely to contribute to ___ in the image.
Answer
-
quantum mottle
-
motion
-
high spatial resolution
-
excessive OID
Question 83
Question
Using the 15% rule changes
Answer
-
distortion
-
IR exposure
-
remnant beam contrast
Question 84
Question
A radiograph of the forearm is produced using 4 mAs at 65 kVp. What kVp would be required to double the exposure? round up to whole number
[blank_start]75[blank_end] kVp
Question 85
Question
As OID increases, size distortion will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 86
Question
If the signal is 16 and the noise is 4 on a hypothetical radiographic image, what is the Signal to Noise ratio?
Question 87
Question
Which of the following is equivalent to low penumbra?
Question 88
Question
Penumbra decreases as
Answer
-
SID increases
-
OID decreases
-
focal spot decreases
-
all of the above
Question 89
Question
If the mA is set at 300 and the total mAs produced is 15 mAs, the exposure time must be:
Answer
-
0.5 sec
-
0.05 sec
-
0.2 sec
-
2 sec
Question 90
Question
Size distortion in radiography can be ___ only.
Answer
-
magnification
-
foreshortening
-
elongation
-
minimization
Question 91
Question
What factor influences the average energy of photons reaching the IR?
Answer
-
mAs
-
kVp
-
mA
-
focal spot size
Question 92
Question
If 20 mAs at 100 kVp is used and the IR exposure needs to be doubled without altering the remnant beam contrast scale, determine the new technique that should be used.
new kVp: [blank_start]100[blank_end]
new mAs: [blank_start]40[blank_end]
Question 93
Question
which of the following are causes of scatter radiation
Answer
-
high kVp
-
low kVp
-
large part thickness
-
part density
-
large field sizes
-
small field sizes
-
large mAs
-
small mAs
Question 94
Question
what is the purpose of a grid
Answer
-
to absorb low energy photons before they reach the IR
-
to absorb scatter before it reaches IR
-
to reflect scatter away from IR
-
to absorb primary radiation before it reaches IR
Question 95
Question
match the type of grid with its definition
[blank_start]Stationary[blank_end] grid: attaches to cassette
[blank_start]Reciprocating[blank_end] grid: moves grid back and forth
[blank_start]Oscillating[blank_end] grid: move grid plate in circular motion
The [blank_start]reciprocating[blank_end] and [blank_start]oscillating[blank_end] grids make up the Potter-Bucky Diaphragm aka the [blank_start]bucky[blank_end]
Answer
-
Stationary
-
Reciprocating
-
Oscillating
-
reciprocating
-
oscillating
-
bucky
Question 96
Question
the [blank_start]higher[blank_end] the grid ratio, the more effective the grid is at removing scatter
Question 97
Question
We used 10 mAs at 78 kV on a KUB. Convert this non-grid technique to a 10:1 grid
[blank_start]50[blank_end] mAs @ [blank_start]78[blank_end] kV
Question 98
Question
Convert 60 mAs at 92 kV using a 8:1 grid to an 5:1 grid.
[blank_start]30[blank_end] mAs @ [blank_start]92[blank_end] kV
Question 99
Question
increased lead content = [blank_start]decreased[blank_end] scatter = contrast scale [blank_start]shortens[blank_end] = image contrast [blank_start]increases[blank_end]
Answer
-
decreased
-
increased
-
same
-
shortens
-
lengthens
-
remains the same
-
increases
-
decreases
-
remains the same
Question 100
Question
grid cut off is due to the absorption of
Answer
-
scatter radiation
-
remnant radiation
-
primary radiation
-
secondary radiation
Question 101
Question
label the type of grid
Answer
-
short axis
-
linear
-
cross-hatch
Question 102
Question
label the type of grid
Question 103
Question
What is the grid radius based on?
Answer
-
SID
-
OID
-
Tube angle
-
Object thickness
Question 104
Question
fill out the table and GCF equation, include x in multiplier.
Answer
-
No grid
-
x1
-
5:1
-
x2
-
6:1
-
x3
-
8:1
-
x4
-
10:1 or 12:1
-
x5
-
16:1
-
x6
-
mAs2
-
mAs1
-
GCF2
-
GCF1
Question 105
Question
What causes the moire effect?
Answer
-
the grid lines are parallel to scan lines of processor
-
the grid lines are perp to scan lines of processor
-
misalignment of the grid and beam
-
incorrect SID used
Question 106
Question
This image is an example of
Question 107
Question
Match the image with its grid error
Question 108
Question
The air gap technique is when OID is [blank_start]increases[blank_end], so scatter on the IR [blank_start]decreases[blank_end]
Answer
-
increased
-
decreased
-
remains the same
-
increases
-
decreases
-
remains the same
Question 109
Question
When generator power and rectification increase
Answer
-
quantity of current increases
-
quantity of current decreases
-
quality of current increases
-
quality of current decreases
-
current is not affected
Question 110
Question
high frequency generators [blank_start]reduce[blank_end] the ripple of the wave form
with each increase in generator power, the exposure rate of the x-ray beam [blank_start]increases[blank_end]
These changes [blank_start]increase[blank_end] the effective mA, which allows for a [blank_start]reduction[blank_end] in technique
High power generators also [blank_start]increase[blank_end] the average kV
Generally, the set mAs could be [blank_start]cut in half[blank_end] when changing from a single phase to a 3 phase/HF
Answer
-
reduce
-
increase
-
don't change
-
increases
-
decreases
-
remains the same
-
increase
-
decrease
-
do not change
-
decrease
-
increase
-
no change
-
increase
-
decrease
-
do not change
-
cut in half
-
doubled
-
unchanged
Question 111
Question
generator power [blank_start]lengthens[blank_end] remnant contrast scale
generator power [blank_start]does not affect[blank_end] spatial resolution
generator power [blank_start]does not affect[blank_end] size distortion
generator power [blank_start]does not affect[blank_end] shape distortion
Answer
-
does not affect
-
shortens
-
lengthens
-
does not affect
-
increases
-
decreases
-
does not affect
-
increases
-
decreases
-
does not affect
-
increases
-
decreases
Question 112
Question
which of the following are considered inherent filtration
Question 113
Question
which of the following are considered added filtration
Question 114
Question
The minimum total filtration equivalent is [blank_start]2.5[blank_end] mm Al
- [blank_start]1.5[blank_end] mm of inherent filtration
- [blank_start]1[blank_end] mm added filtration
Question 115
Question
What is the unit for directly measuring x-ray beam penetration
Answer
-
Half-Value Layer
-
Half-Vacancy Layer
-
mm Al
-
mm Hg
-
mm Pb
Question 116
Question
A Potter-Bucky diaphragm does not move
Question 117
Question
Higher frequency grids have [blank_start]thinner[blank_end] lead strips
Answer
-
thinner
-
thicker
-
no difference in
Question 118
Question
What is the most critical factor in a grid's efficiency?
Answer
-
grid frequency
-
grid ratio
-
lead content of grids
-
grid dimension
Question 119
Question
Grid radius is the distance between the grid and the [blank_start]convergence line[blank_end].
Question 120
Question
Match the term with its definition
[blank_start]Grid ratio[blank_end] : the proportion of the height to the width of the interspaced between the lead strips
[blank_start]Lead Content of Grid[blank_end] : mass per unit area, as this increases so does the removal of scatter
[blank_start]Grid Radius[blank_end]: the distance between the grid and the convergent line or point aka the correct SID
[blank_start]Grid Frequency[blank_end]: the number of lead strips counted per inch/cm
Answer
-
Grid ratio
-
Lead Content of Grid
-
Grid Radius
-
Grid Frequency
Question 121
Question
What is the purpose of a compensating filter?
Question 122
Question
Which of the follower are the purposes for field size limitation
Answer
-
minimize radiation exposure to the tech
-
minimize radiation exposure to the patient
-
change subject contrast in the remnant beam
-
preserve subject contrast in the remnant beam
-
preserve image sharpness
Question 123
Question
increased collimation (smaller size) = [blank_start]decreased[blank_end] scatter = [blank_start]shortened[blank_end] contrast scale
increased collimation = [blank_start]decreased[blank_end] IR exposure
Answer
-
increased
-
decreased
-
same amount of
-
lengthened
-
shortened
-
same
-
decreased
-
increased
-
same
Question 124
Question
match the types of field size limitations
Question 125
Question
The purpose of PBL is
Answer
-
to allow techs to not worry about collimation
-
ensure the field size is never larger than cassette
-
allow for easy collimation
-
automatically set field size
Question 126
Question 127
Question
Positive Contrast Agents
-[blank_start]Radiopaque[blank_end] ([blank_start]white[blank_end])
-Based upon [blank_start]Barium or Iodine[blank_end]
Negative Contrast Agents
-[blank_start]Radiolucent[blank_end] ([blank_start]black[blank_end])
-Are all [blank_start]gasses[blank_end]
Answer
-
Radiopaque
-
Radiolucent
-
white
-
black
-
Barium or Iodine
-
Barium or Calcium
-
Lead or Tungsten
-
Tungsten or Molybdenum
-
Calcium or Iodine
-
Barium or Tungsten
-
Radiopaque
-
Radiolucent
-
white
-
black
-
gasses
-
water
-
tissues
-
bones
-
barium
-
iodine
Question 128
Question
Additive diseases = [blank_start]increased radiopacity of tissue[blank_end] = [blank_start]increased[blank_end] technique
Destructive diseases = [blank_start]increased radiolucency of tissue[blank_end] = [blank_start]decreased[blank_end] technique
Answer
-
increased radiopacity of tissue
-
increased radiolucency of tissue
-
decreased radiopacity of tissue
-
decreased radiolucency of tissue
-
increased
-
decreased
-
same
-
increased radiopacity of tissue
-
increased radiolucency of tissue
-
decreased radiopacity of tissue
-
decreased radiolucency of tissue
-
increased
-
decreased
-
same
Question 129
Question
which of the following are additive diseases
Answer
-
Emphysema
-
Pleural Effusion
-
Edema
-
Osteoporosis
-
Pneumothorax
-
Pneumonia
-
Cardiomegaly
-
Cancer
-
increase fluid, bone or metal content
-
increase gas or fat, decrease normal fluid or bone
Question 130
Question
which of the following are destructive diseases
Answer
-
Emphysema
-
Pleural Effusion
-
Edema
-
Osteoporosis
-
Pneumothorax
-
Pneumonia
-
Cardiomegaly
-
Cancer
-
increase fluid, bone or metal content
-
increase gas or fat, decrease normal fluid or bone
Question 131
Question
For soft tissue techniques, kVp is
Question 132
Question
For patients with a fiberglass cast/splint, the technique may be altered by
Answer
-
an increase in mAs by 50%
-
an increase in mAs by 20%
-
an increase in mAs by 8%
-
an increase in kVp by 50%
-
an increase in kVp by 20%
-
an increase in kVp by 8%
Question 133
Question
the longer the SID, the [blank_start]sharper[blank_end] the detail
the longer the SID, the [blank_start]less[blank_end] magnification
the longer the SID, the [blank_start]no effect on[blank_end] shape distortion
Answer
-
sharper
-
fuzzier
-
no effect on
-
more
-
less
-
no effect on
-
more
-
less
-
no effect on
Question 134
Question
fill in the formula for the inverse square law
Question 135
Question
fill in the formula for the exposure maintenance law
Question 136
Question
Increased OID = [blank_start]Decreased[blank_end] Scatter = [blank_start]shorter[blank_end] contrast scale
Increased OID = [blank_start]decreased[blank_end] scatter = [blank_start]decreased[blank_end] IR exposure
Increased OID = [blank_start]Decreased[blank_end] sharpness
Increased OID = [blank_start]Increased[blank_end] magnification
Increased OID = [blank_start]No effect on[blank_end] shape distortion
Answer
-
decreased
-
increased
-
no effect on
-
shorter
-
longer
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
-
decreased
-
increased
-
no effect on
Question 137
Question
If 40 mAs at 76 kVp is used and the amount of exposure needs to be cut in half, determine the new technique that should be used if you want to use the 15% rule. round to 1 decimal place if necessary
[blank_start]64.6[blank_end] kVp @ [blank_start]40[blank_end] mAs
Question 138
Question
If 80mAs at 70kVp is used and the IR exposure needs to be doubled and the remnant beam scale of contrast lengthened, determine the new technique that should be used. round to one decimal if necessary
new kVp: [blank_start]80.5[blank_end]
new mAs: [blank_start]80[blank_end]
Question 139
Question
A radiograph of the forearm is produced using 4 mAs at 65 kVp. round to 1 decimal place if necessary
new kVp to double IR exposure: [blank_start]74.8[blank_end]
new kVp to halve IR exposure: [blank_start]55.3[blank_end]
Question 140
Question
Volume of the irradiated tissue increases as
Question 141
Question
The half-value layer (HVL) is used to measure
Question 142
Question
Photon absorption within the body occurs as the result of
Question 143
Question
The [blank_start]____[blank_end] the photon energy, the [blank_start]____[blank_end] the percentage of interactions that will be Compton interactions.
Answer
-
higher
-
lower
-
higher
-
lower
-
unaffected
Question 144
Question
If OID increases, the remnant beam contrast scale [blank_start]_________________[blank_end].
Answer
-
lengthens
-
shortens
-
is unaffected
Question 145
Question
The computer distinguishes the anatomy of interest from background exposure outside collimated field during:
Question 146
Question
How dark or light a digital image appears on a display monitor is known as
Answer
-
contrast resolution
-
brightness
-
spatial resolution
-
density
Question 147
Question
In generic digital imaging terms, the average brightness level of the image is called the:
Answer
-
window level
-
window width
-
density
-
scale of contrast
Question 148
Question
In a digital image histogram graph, the vertical height of any data point along the plotted curve (measured against the y axis of the graph), indicates the:
Question 149
Question
A barium study projection will likely require the computer to locate both Smin and Smax.
Question 150
Question
On an image histogram, an unusual spike to the left of the main bell-shaped lobe of the curve most likely represents:
Answer
-
osteoporosis
-
background exposure
-
a large metallic object
-
air in the lungs
Question 151
Question
Dependent on the manufacturer’s algorithm, a value of interest (VOI) can accentuate a certain tissue thickness or bone.
Question 152
Question
The larger the number of pixels in an image matrix, the:
Answer
-
shorter the acquisition time
-
greater the resolution of the image
-
greater the number of gray shades in the image
-
all answers are correct
Question 153
Question
Which of the following is an example of a Field Uniformity Correction?
Answer
-
creation of a best fit line for the raw detector data
-
a flattening of the histogram curve as needed
-
application of a kernel across the matrix
-
evening out of brightness to account for irregularities in the hardware of the IT across the image field
Question 154
Question
Many imaging systems present the histogram with an averaged gray scale curve overlying the histogram. The steeper the slope:
Question 155
Question
Digital radiography images that are underexposed have a high SNR.
Question 156
Question
The development of which of the following enabled DR systems to use a flat-shaped receptor plate:
Question 157
Question
In digital imaging processing, what is segmentation?
Answer
-
a graph plotting signal intensity for each pixel
-
identification of the exposed area of the IR
-
the anatomy on which the exam is performed
-
a look up table
Question 158
Question
Generally speaking, the smaller the pixels, the higher the spatial resolution.
Question 159
Question
Which of the following terms, all contractions, most accurately describes the individual hardware elements of a DR flat panel receptor that detect radiation in the remnant x-ray beam?
Answer
-
the dexel
-
the pixel
-
the voxel
-
the rexel
Question 160
Question
The selection of a narrow window width will:
Answer
-
decrease the size of each pixel
-
decrease overall brightness of the image
-
increase contrast in the image
-
increase magnification of the central portion of the image
Question 161
Question
Which of the following modalities create images with the highest spatial resolution?
Answer
-
ct
-
nuclear medicine
-
ultrasound
-
digital radiography
Question 162
Question
Adjustment of the window width of the image will change
Answer
-
the brightness of the image
-
the level of IR exposure
-
the spatial resolution of the image
-
the scale of contrast of the image
Question 163
Question
The smaller the pixels, the lower the spatial resolution.
Question 164
Question
For digital imaging, each image displayed consists of a discrete number of rows and columns of picture elements called the:
Answer
-
signal plate
-
display table
-
matrix
-
detector array
Question 165
Question
In digital radiography, image contrast displayed on the monitor is controlled exclusively by kilovoltage.
Question 166
Question
The larger the number of pixels in an image matrix, the:
Answer
-
greater the number of gray shades in the image
-
shorter the acquisition time
-
greater the spatial resolution in the image
-
all answers are correct
-
no answers are correct
Question 167
Question
A system with a high S/N ratio displays:
Answer
-
low noise and high signal
-
low signal and high noise
-
equal amounts of signal and noise
-
no signal and no noise
Question 168
Question
Scatter radiation adds exposure to the image receptor.
Question 169
Question
Which factor influences the average energy of photons reaching the IR?
Answer
-
kVp
-
mA
-
time
-
focal spot size
Question 170
Question
As kVp decreases, spatial resolution will _________________.
Answer
-
increase
-
decrease
-
remain unchanged
Question 171
Question
In the diagnostic range, the majority of primary x-rays produced are:
Answer
-
characteristic
-
photoelectric
-
brems
-
compton
Question 172
Question
As kVp increases,
Question 173
Question
The amount of scatter reaching the image receptor is affected by all of the following EXCEPT:
Answer
-
SID
-
Amount of Collimation
-
kVp
-
grid ratio
Question 174
Question
The larger the number of pixels in an image matrix, the:
Question 175
Question
The higher the DQE, the higher the dose required to get an optimal image.
Question 176
Question
As the lead content of a grid increases,
Answer
-
image receptor exposure increases
-
the contrast decreases
-
The ability of the grid to remove scatter increases.
-
The ability of the grid to remove scatter decreases
Question 177
Question
In production of a radiograph, which of the following is considered the controlling factor of beam penetration and remnant beam scale of contrast?
Question 178
Question
The principal controlling factor of radiographic image receptor exposure is:
Answer
-
kVp
-
mAs
-
focal spot size
-
tube angulation
Question 179
Question
Image level parameters (window level) control image
Answer
-
brightness
-
contrast
-
density
-
darkness
Question 180
Question
All of the following are true of pixel pitch except:
Answer
-
it is measured from center-center of pixels
-
it is affected by changes in the matrix size
-
it is determined by the size of the dels in flat panel systems
-
it is not related to spatial resolution
Question 181
Question
Which of the following is NOT a component of a detector element?
Answer
-
capacitor
-
charged coupled device
-
semiconductor layer
-
thin film transistor
Question 182
Question
A phosphor screen or scintillator
Answer
-
fluoresces when exposed to light
-
fluoresces when exposed to xrays
-
ionizes selenium when exposed to x-rays
-
ionizes selenium when exposed to light
Question 183
Question
The percentage of the surface area of a flat panel detector element devoted to the semiconductor layer is:
Answer
-
DQE
-
Fill factor
-
detector ratio
-
capacitance
Question 184
Question
The purpose of the ADC is to convert the _______ signal to a _________ signal.
Answer
-
analog; digital
-
digital; analog
-
analog; detector
-
acquired; detected
-
detector; digital
Question 185
Question
Digital radiography images that are underexposed can show quantum mottle.
Question 186
Question
In a DR flat panel image receptor, which electronic devices directly account for the plate’s ability to store a latent image until it is sent into the computer?
Answer
-
capacitor
-
tfts
-
data lines
-
gate lines
-
semiconductor layer
Question 187
Question
An x-ray unit with a Modulation Transfer Function (MTF) of 0.2 is said to have perfect contrast resolution.
Question 188
Question
What is the size of each pixel for an image with a 1024x1024 matrix using a 30x30 cm field of view:
Question 189
Question
The ratio of SID to SOD is the primary control for:
Question 190
Question
In a DR flat panel image receptor, most of the square area of each del is dedicated to what portion of the detector element:
Answer
-
the semiconductor layer
-
the gate lines
-
the capacitors
-
the tfts
Question 191
Question
The ionization chamber AEC is located behind the IR on the wall Bucky.
Question 192
Question
As matrix size increases, the image's spatial resolution increases.
Question 193
Question
Direct conversion digital imaging systems include which of the following?
Answer
-
CCD
-
PSP
-
amorphous Se tft
-
amorphous Si tft
Question 194
Question
A CMOS imaging system would have
Answer
-
microlens optics
-
charged couple arrays
-
amorphous selenium
-
a helium red laser
Question 195
Question
Which of the following is NOT one of the three steps to digitization of an analog image?
Answer
-
Sampling
-
Quantization
-
Scanning
-
Initiation
Question 196
Question
The maximum range of pixel values a hardware device can store is:
Answer
-
voxel depth
-
bit depth
-
gray scale
-
gray level
Question 197
Question
A radiograph is made using the following factors:
100 mA 8:1 grid
1.0 Sec 8x10 collimated field size
70 kVp large focal spot
50" SID 3mm Al
2" OID suspended respiration
In each of the following, if only the proposed change is made, how will spatial resolution be affected?
- Use of a 10x12 field size: [blank_start]no change[blank_end]
- Use of no grid: [blank_start]no change[blank_end]
- Use 200 mA: [blank_start]no change[blank_end]
- Use 4" OID: [blank_start]decrease[blank_end]
- Use small focal spot: [blank_start]increase[blank_end]
- Use 4mm Al: [blank_start]no change[blank_end]
-Use 4 seconds: [blank_start]decrease[blank_end]
Answer
-
no change
-
increase
-
decrease
-
no change
-
increase
-
decrease
-
no change
-
increase
-
decrease
-
decrease
-
increase
-
no change
-
increase
-
decrease
-
no change
-
no change
-
increase
-
decrease
-
decrease
-
increase
-
no change
Question 198
Question
A radiograph is made using the following factors:
100 mA 8:1 grid
1.0 Sec 8x10 collimated field size
70 kVp large focal spot
50" SID 3mm Al
2" OID suspended respiration
In each of the following, if only the proposed change is made, how will IR exposure be affected?
- Use of a 10x12 field size: [blank_start]increased[blank_end]
- Use of no grid: [blank_start]increased[blank_end]
- Use 200 mA: [blank_start]increased[blank_end]
- Use 2.0 seconds: [blank_start]increased[blank_end]
- Use 4" OID: [blank_start]decreased[blank_end]
- Use 60" SID: [blank_start]decreased[blank_end]- Use 90 kVp: [blank_start]increased[blank_end]
Answer
-
increased
-
decreased
-
no change
-
increased
-
decreased
-
no change
-
increased
-
decreased
-
no change
-
increased
-
decreased
-
no change
-
decreased
-
increased
-
no change
-
decreased
-
increased
-
no change
-
increased
-
decreased
-
no change
Question 199
Question
A radiograph is made using the following factors:
500 mA
2.5 sec
130 kVp
36" SID
6" OID
2000x1000 matrix size
Window Width = 800
No grid
5x8 collimated field size
small focal spot
3mm Al
0.1mm pixel size
Window Level = 65
Changes are proposed below. In each of the following, if only the proposed change is made, how will spatial resolution be affected?
- Collimate to 10x10; [blank_start]n/a[blank_end]
- Use a grid; [blank_start]n/a[blank_end]
- Use 72" SID; [blank_start]increase[blank_end]
- Use 1" OID; [blank_start]increase[blank_end]
-Window level = 165; [blank_start]n/a[blank_end]
- Use 3600x3600 matrix size; [blank_start]increase[blank_end]
- Use 0.1cm pixel size; [blank_start]decrease[blank_end]
Answer
-
n/a
-
increase
-
decrease
-
n/a
-
increase
-
decrease
-
increase
-
decrease
-
n/a
-
increase
-
decrease
-
n/a
-
n/a
-
increase
-
decrease
-
increase
-
decrease
-
n/a
-
decrease
-
increase
-
n/a