ST Monitoring Quiz

Beschreibung

ST monitoring via Philips monitor
Natalie Tredway
Quiz von Natalie Tredway, aktualisiert more than 1 year ago
Natalie Tredway
Erstellt von Natalie Tredway vor mehr als 9 Jahre
278
1

Zusammenfassung der Ressource

Frage 1

Frage
ST analysis can be performed on the patient with 1:1 ventricular pacing.
Antworten
  • True
  • False

Frage 2

Frage
The ST segment is:
Antworten
  • Normally elevated in all leads
  • Normally isoelectric in all leads
  • Normally depressed in all leads
  • Damaged and ischemia causes the cells of the myocardium to become more or less excitable.

Frage 3

Frage
ST segment measurements are displayed and trended in:
Antworten
  • mmHg
  • mm
  • kpal

Frage 4

Frage
The ST Leads that are being monitored depend on: (choose all that apply)
Antworten
  • The leads available based on the lead set used and electrodes applied.
  • The Lead Placement selected - Standard or EASI
  • The number of leads displayed on the Main Screen.
  • The leads setup for monitoring.

Frage 5

Frage
When ST Analysis is turned off:
Antworten
  • No ST analysis or arrhythmia analysis is performed.
  • No ST analysis is performed but arrhythmia analysis continues.

Frage 6

Frage
To turn off individual leads of ST:
Antworten
  • Change to another screen
  • In the ST Analysis menu, select Setup ST Leads.
  • In the ECG menu, select New Lead Setup.
  • In the ST Analysis menu, select, select ST Analysis.

Frage 7

Frage
Selecting Alarms in the ST Analysis menu:
Antworten
  • Turns on/off all ST alarms.
  • Allows you turn off individual ST alarms.
  • Silences ST alarms for 5 minutes.
  • Allows you to select single-lead or multi-lead alarms.

Frage 8

Frage
Multi-lead alarms will cause an alarm only when:
Antworten
  • All monitored ST leads exceed the limits set for 15 seconds.
  • Three or more leads exceed the limits set for 15 seconds.
  • Two or more leads exceed the limits set for 1 minute.
  • Three or more leads exceed the limits set for 1 minute.

Frage 9

Frage
ST alarms set using Auto Limits Wide: (choose all that apply)
Antworten
  • Changes the alarm limits based on the last 5 minutes of data.
  • Change as the patient’s condition changes
  • Changes the alarm limits based on the last 1 minute of data.
  • Will not change until you change them manually or use AutoLimits again.
  • Will not set alarm limits in pathological range.
Zusammenfassung anzeigen Zusammenfassung ausblenden

ähnlicher Inhalt

Französisch für Anfänger
JohannesK
Elektrische Spannung
Peter Kasebacher
Deutsch Einstufungstest Niveau A2.1
SprachschuleAktiv
Hardware- und Gerätetechnik
DFairy
Kunststoffe - Einführung
Fabian B.
Biologie Abiturthemen 2016
katharina.boerge
Machst du auch diese 10 typischen Fehler auf Deutsch?
Dilyana Hunley
Grundlagen der BWL
Robert Ris
EC angewandte Sozialpsychologie Uni Wien
Alina S.
Vetie - Histo & Embryo II 2017
Fioras Hu
Biwi05 Fragenkatalog
Robin Pietsch