Questão 1
Questão
Gas Chromatography (Absorption Type)
Sample gas moves through a [blank_start]separator[blank_end] column where the differing sizes of molecules for the individual gases make each gas exit at a different [blank_start]rate[blank_end],
Questão 2
Questão
Advantages: Analyzes [blank_start]many gases[blank_end] at once. Very accurate
Disadvantages: Requires more [blank_start]time[blank_end] than other analyzers.
Affected by: [blank_start]H2O[blank_end] and [blank_start]CO2[blank_end]
Measures: [blank_start]Ne[blank_end], [blank_start]O2[blank_end], [blank_start]N2[blank_end], [blank_start]CO[blank_end]
Responda
-
CO
-
N2
-
O2
-
Ne
-
CO2
-
H2O
-
time
-
many gases
Questão 3
Questão
Mass Spectometry
Function: Sample is drawn in by vacuum to the [blank_start]ionization[blank_end] chamber. Individual gases are [blank_start]separated[blank_end] and analyzed.
Measures: Gas percentages
Advantages: [blank_start]Rapid[blank_end] Response time, breath by breath analysis, multiple gas analysis
Can analyze: [blank_start]He[blank_end], [blank_start]N2[blank_end], [blank_start]O2[blank_end], [blank_start]CO2[blank_end]
Disadvantages: Expensive, large, high maintenance
Most Suitable for: analysis of [blank_start]several[blank_end] patient at once rather than individual testing.
Responda
-
ionization
-
separated
-
Rapid
-
He
-
N2
-
O2
-
CO2
-
several
Questão 4
Questão
CO2/CO/CH4
AKA: Infrared Absorption Analyzer
Quick response time is suitable for [blank_start]DLCO[blank_end]
Calibrate: to room air ([blank_start]0[blank_end]%) and known concentration ([blank_start].3[blank_end]%) CO then re-zero and use [blank_start]gain[blank_end] to match read out to [blank_start]known[blank_end] concentration.
Changes in altitude do [blank_start]NOT[blank_end] affect readings.
Questão 5
Questão
Helium analyzer
Also Known as: [blank_start]Thermal[blank_end] Conductivity or [blank_start]Wheatstone[blank_end] Bridge
Measures: Gas [blank_start]percentage[blank_end]
Calibration: To room air ([blank_start]0[blank_end]%) and known ([blank_start]10[blank_end]%) then use gain to adjust to [blank_start]known[blank_end] amount.
Do not use with [blank_start]flammable[blank_end] gases.
Used with Helium dilution tests.
Responda
-
Thermal
-
Wheatstone
-
percentage
-
0
-
10
-
known
-
flammable
Questão 6
Questão
Nitrogen analyzer
Also known as: [blank_start]Geissler[blank_end] Tube Ionizer
Sample is pumped into [blank_start]ionization[blank_end] chamber where light is emitted and monitored by a phototube. Intensity of light is directly [blank_start]proportional[blank_end] to N2%
Measures gas percentage
Capable of breath by breath analysis
Calibration: To zero with [blank_start]100[blank_end]% O2 and a [blank_start]known[blank_end] concentration.
Responda
-
Geissler
-
ionization
-
proportional
-
100
-
known
Questão 7
Questão
O2 Analyzers
Galvanic [blank_start]Fuel[blank_end] Cell Analyzer: Measures partial [blank_start]pressure[blank_end] and displays [blank_start]FiO2[blank_end] as %.
If unable to calibrate you should replace the [blank_start]fuel[blank_end] cell.
Calibration during exercise testing: Low [blank_start]12[blank_end]-[blank_start]14[blank_end]% and high [blank_start]21[blank_end]%
Poloragraphic Analyzer
Similar to galvanic but uses a [blank_start]battery[blank_end]
Analogous to the [blank_start]Clark[blank_end] electrode
Calibration is to [blank_start]21[blank_end]% and [blank_start]100[blank_end]%
Responda
-
Fuel
-
pressure
-
FiO2
-
fuel
-
12
-
14
-
21
-
21
-
100
-
Clark
-
battery
Questão 8
Questão
Gas & Water Absorbers
CO2 Absorber: AKA Sodium [blank_start]Hydroxide[blank_end], NaOH (soda [blank_start]lime[blank_end] or soda-sorb), Baralyme
Required for [blank_start]rebreathing[blank_end] tests (FRC by He Dilution, DLCO-RB)
Replace scrubber: When [blank_start]exhausted[blank_end], when [blank_start]color[blank_end] changes, or weekly.
CO2 Scrubber produce [blank_start]water[blank_end] as a by product and should be placed [blank_start]upstream[blank_end] from water absorbers.
Responda
-
lime
-
Hydroxide
-
rebreathing
-
exhausted
-
color
-
water
-
upstream
Questão 9
Questão
Gas & Water Absorbers
Water Absorber ([blank_start]Dessicant[blank_end])
Drierite ([blank_start]Calcium[blank_end] Sulfate)
Removes moisture as patient breathes through system.
Proper placement of absorbers:
1st: [blank_start]CO2[blank_end] Absorber
2nd: [blank_start]H2O[blank_end] Absrober
3rd: Anaylzer
Permapure (fafion Tubing)
Conditions gas to [blank_start]equilibrate[blank_end] with ambient humidity
Allows for accurate correction for humidity in a gas sample.
Responda
-
Calcium
-
Dessicant
-
CO2
-
H2O
-
equilibrate